TEST PLAN

CDSI-R-001 Test Plan

Test plan published before any data exist. Open for comment: protocols@cdsi.click

ReportCDSI-R-001StatusDraft plan v0.1, not yet runMethodCDSI-001 v1.1

Orthogonal two-lab off-gas characterization of the Divine Tribe V5, run at its recommended-use setting

Status: draft for public comment. Nothing has been run. This plan is published before any data exists so the method can be criticized before money is spent.

In print: We’re Testing My Own Vaporizer in Public, and the Plan Is Published Before Anything Runs (Marijuana Union, November 2026).

1. Purpose

Measure what comes off one vaporizer’s heater during one ordinary session, using two independent labs that use two different sampling methods (orthogonal testing). Agreement between the two cross-validates the result. Disagreement is itself a finding about method, and gets published the same way.

This run characterizes the device only. No cannabis or other controlled substance is used. The heater runs empty (or with an inert load, see §4.3), so whatever is measured comes from the hardware.

2. Conflict of interest (cover-page disclosure)

The first device tested is made by Divine Tribe / Nice Dreamz LLC, owned by CDSI’s founder, Matthew Macosko, who directs this test. Disclosed under CDSI COI Policy Article VIII. Any industry ties of either lab will be disclosed on the cover as well. Other brands follow in later reports under the same plan.

3. Device

  • Make/model: Divine Tribe V5 (concentrate vaporizer)
  • Units: 3 from regular production stock, serial numbers logged and photographed
  • Condition: new, no pre-conditioning beyond the manufacturer’s stated first-use instructions (stated in the report)

4. The run (recommended-use cycle)

4.1 Setpoint: 500 °F (260 °C). This is a high-heat session: the top of the range Divine Tribe recommends to users, chosen deliberately so the result reflects a realistic worst case, not a gentle one.

4.2 Session (vacuum draw while dry-firing): the heater is dry-fired at setpoint for 60 seconds while a vacuum pump pulls air through the device the entire minute, the same way a person inhaling through it would. Air enters the device’s own air path, passes the hot heater, and exits the mouthpiece into the sampling train, so the sample is what a user’s lungs would receive. - Draw flow: [TBD — target a human inhalation rate through this device; to be agreed with both labs and stated in the report]. - Sampling tubes each have a maximum flow, often lower than a human draw. If so, the main vacuum carries the full session flow and each tube pulls a measured side-stream from it, and the report states both flows so the numbers can be scaled back to one session. - Lab B’s puffing machine runs the same 60-second, 500 °F, continuous-draw session (not its default puff regime), so both labs test the same thing.

4.2a Draw-through only, never a chamber or oven bake. The device is NOT placed in a heated chamber, oven or enclosure and off-gassed into it. Off-gassing a whole device in a chamber measures everything that outgasses from its body, battery and plastics, at temperatures a user never exposes them to, and is not what a user breathes. CDSI-R-001 measures only the air a user would inhale: pulled in through the device’s own intake, across its own heater, out its mouthpiece, under its own power. Any lab that can only run a chamber method is not used for this report. (Matches CDSI-001 v1.1 §3.1–3.1a, amended 2026-11-12.)

4.3 Load: empty heater. An inert-load arm (e.g., clean glass wool) is a possible addition; comments welcome.

4.4 Sessions per sample: one minute may not collect enough for the labs to measure. Each sampling tube collects N identical sessions (N = [TBD from the labs’ stated minimum mass], provisionally 5), with a cool-down to room temperature between sessions. Every session is a real user session; only the count is chosen for detection.

4.5 Blanks: a room-air blank (same pump, same time, no device) and a field blank (tube opened and capped, no flow) with every sample set. A run is invalid if blanks exceed the lab’s reporting limit for any target.

4.6 Heater temperature check: a thermocouple on the heater logs actual temperature every second, so the report states what the heater really reached, not just the dial setting.

5. Lab A: industrial-hygiene sorbent method (founder-run, on camera)

  • Sampling done by Matt at the CDSI bench, filmed start to finish.
  • Media: DNPH cartridge (carbonyls: formaldehyde, acetaldehyde, acrolein + 13-carbonyl scan), sorbent tube for VOCs (BTEX + scan), filter for metals (ICP).
  • Lab: an independent industrial-hygiene laboratory with published pricing (named in the report once engaged).
  • Chain of custody: labeled, photographed, sealed, shipped cold where required; lab chain-of-custody form copied into the report.

6. Lab B: vape-emissions puffing-machine method (lab-run, founder directing)

  • Lab runs the device on its own puffing machine under its own method (named in the report once engaged; accreditation status stated).
  • Matt is physically present and directs every session on site: setpoint, session length, session count, blanks, same as §4. A lab that will not allow the founder in the room is not used.
  • Same analyte families as Lab A so results can be compared.

7. What gets published (all of it)

  1. This plan, before the run, open for comment
  2. Full video of every Lab A session; Lab B session video if the lab permits
  3. Thermocouple logs (CSV)
  4. Photos of every device serial, tube, label and seal
  5. Chain-of-custody forms
  6. Both labs’ raw reports, unedited
  7. A side-by-side comparison table with each lab’s reporting limits
  8. Plain-language summary

Language rule: nothing is reported as “zero.” Results below a lab’s limit are reported as “below the attribution floor” for that named analyte, at that lab’s stated limit, at 500 °F for this session length (CDSI Lexicon §2A).

8. Open questions for reviewers

  • Is empty-heater the right first arm, or should an inert load be in the first run?
  • Is N = 5 sessions per tube the right trade-off between real use and detectability?
  • Which carbonyl/VOC/metal targets would you add or drop?
  • Flow rate: match a standard puffing regime, or measured real-user draw?

9. Budget (estimate, to be replaced by real quotes)

  • Lab A: ~$1,000–1,700 for 3 units + blanks (2021 published prices, to be re-quoted)
  • Lab B: quote pending
  • Travel to Lab B for on-site attendance: [TBD]
  • Thermocouple/logger: <$150